ELISA Reader
Light Source
- 8 tungsten lamps, automatically underpowered when in stand-by position
Optical Filters
- 2 filters (450, 630 nm); 2 more on request (max 4 filters, 400-700nm)
Detectors
- 8 silicon photodiodes
Reading System
- 8 independent channels, mono or bichromatic reading
Inaccuracy
- ± 1% from 0.000 to 1.500 OD
- ± 2% from 1.500 to 3.000 OD
Reading Time
- 16s to perform a monochromatic reading on a full plate; 31s for bichromatic reading
Cuvettes
- 96 well plate
Plate Shaking
- Programmable by keyboard
LCD-Display
- 2 lines with 16 columns
Printer
- 40 columns on thermal paper
Keypad
- 17 buttons
Interface
- RS232 serial port
Curves
- Curve direct plotting: point to point, linear regression and 4 parameters (log-logit)
Results
- by strip of 8 or 12 wells (X or Y)
Calculation Mode
- OD, direct Cut-Off or calculated, single calibrator, programmable curves from 2 to 12 points
Memory Capacity
- 170 tests
Calendar/Timer
- Associates time with the results; timer function
Power requirements
- 230/115 VAC, 50-60 Hz, 110 W
Measurements
- 42 x 34 x 15 cm
Weight
- 8.3 Kg
Optional Software
- Plate Reader management software “ACTION” for PC